Applies to
FTB-1V2/PRO, FTB-2/PRO & FTB-4-PRO
Introduction
The FTB platform is stuck detecting modules in ToolBox procedures
Procedure to follow
Procedure 1
- Shut down the platform
- Remove the module, charger and battery
- After 10 seconds, insert the battery and connect the charger if needed
- With the module still ejected, power on the platform
- Launch Toolbox if its not done automatically. No module detected should be displayed
- Shut down the platform again
- Insert the module
- Power on the platform and validate if the module is recognized and application can be launched
Procedure 2
- Reboot the unit.
- If you have another unit available, swap the complete backplane by unscrewing the 4 screws at the back. Identify if the problem follows the front or the back of the unit.
- (a) If the problem follows the back, it will need to come in for repair. (b) If the problem stays with the front of the unit, try doing a reset/recovery.
- If problem is still present after a reset, unit will need to come in for repair.
Note - Removing the battery and power supply can resolve the issue.
Should you still have issues, please contact the technical support group of your region.