MaxTester Battery level shows greater than 100%


Applies to

MAX-700B/C/D series, MAX-FIP, MAX-900 series, MAX-5205

Introduction

MaxTester battery power level is showing a higher number than 100% while the Battery indicator displays X.

Procedure to follow

  • This can happen when you remove a battery that was discharged and replace it with a battery that is fully charged. 
  • This will not prevent use of the instrument and the instrument will self calibrate during the next battery cycles.
  • You can also choose to allow the unit to fully discharge and then to recharge, forcing a battery calibration